7 troubleshooting, 1 general errors, Troubleshooting 7.1 – Eppendorf Rolling Cabinet User Manual
Page 23: General errors

Troubleshooting
Rolling Cabinet
English (EN)
23
7
Troubleshooting
7.1
General errors
Only rolling cabinet with transformer:
If the suggested troubleshooting measures fail repeatedly, please contact your
local Eppendorf partner. You can find the contact addresses on the Internet at
www.eppendorf.com/worldwide.
Symptom/
message
Cause
Remedy
The centrifuge is
not functioning
properly.
• No mains power connection.
Check the rolling cabinet/
centrifuge connecting cable.
Check the mains power
connection.
• The rolling cabinet power
switch is not switched on.
Switch the rolling cabinet
power switch on.
• The overcurrent protective
switch has been triggered.
Resetting the overcurrent
protective switch:
1. Switch off the centrifuge power
switch.
2. Switch off the rolling cabinet
power switch.
3. Wait at least 20 s. Switch the
rolling cabinet back on.
4. Switch the centrifuge back on.
The rolling cabinet with
transformer is operational again.
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Mastercycler nexus (118 pages)
- Mastercycler nexus (142 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (13 pages)