7

Troubleshooting

7.1

General errors

Only rolling cabinet with transformer:

If the suggested troubleshooting measures fail repeatedly, please contact your
local Eppendorf partner. You can find the contact addresses on the Internet at
www.eppendorf.com/worldwide.

Symptom/
message

Cause

Remedy

The centrifuge is
not functioning
properly.

• No mains power connection.

Check the rolling cabinet/
centrifuge connecting cable.

Check the mains power
connection.

• The rolling cabinet power

switch is not switched on.

Switch the rolling cabinet
power switch on.

• The overcurrent protective

switch has been triggered.

Resetting the overcurrent
protective switch:

1. Switch off the centrifuge power

switch.

2. Switch off the rolling cabinet

power switch.

3. Wait at least 20 s. Switch the

rolling cabinet back on.

4. Switch the centrifuge back on.
The rolling cabinet with
transformer is operational again.