6 maintenance, 1 cleaning and disinfection, Maintenance 6.1 – Eppendorf Rolling Cabinet User Manual
Page 21: Cleaning and disinfection
Maintenance
Rolling Cabinet
English (EN)
21
6
Maintenance
6.1
Cleaning and disinfection
Clean the accessible surfaces of the rolling cabinet at least once a week or if they
become significantly soiled.
Only rolling cabinets with transformer: Before cleaning/disinfecting, switch off the
rolling cabinet using the mains power switch and pull the power plug.
Cleaning: Use mild cleaning agent.
Disinfection: Use alcohol (ethanol, isopropanol) or alcohol-containing disinfectants.
Wipe the accessible surfaces of the rolling cabinet, accessories and power cable using
a damp cloth.
Thoroughly rinse the non-slip mat after cleaning/disinfection.
Allow the rolling cabinet and accessories to dry thoroughly.
Check the rolling cabinet for damage and corrosion damage.
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning
or disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, clean it
immediately using a mild cleaning agent.
NOTICE! Corrosion due to aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive
polishes.
Do not incubate the accessories in aggressive cleaning agents or disinfectants
for longer periods.
NOTICE! Damage from UV and other high-energy radiation.
Do not use UV, beta, gamma, or any other high-energy radiation for
disinfecting.
Avoid storage in areas with strong UV radiation