4 installation, 1 selecting the location, 2 preparing installation – Eppendorf Rolling Cabinet User Manual
Page 13: Installation 4.1, Selecting the location, Preparing installation

Installation
Rolling Cabinet
English (EN)
13
4
Installation
4.1
Selecting the location
Select the rolling cabinet location according to the following criteria:
• Sturdy, level base at the location
• Distance to adjacent devices and walls: at least 30 cm
• Observe the location and positioning instructions in the centrifuge operating manual.
• Only rolling cabinet with transformer: mains power connection (100 V) according to
name plate. The name plate is located on the rear side of the rolling cabinet.
4.2
Preparing installation
Perform the following steps in the sequence described.
1. Remove the accessories.
2. Remove the cover sheets and top transport protection pads.
3. Remove the packaging foil.
4. Grab the rolling cabinet with your hands from below. Lift the rolling cabinet out of the
packaging board and place it on the ground.
5. Only rolling cabinet with transformer: Allow the rolling cabinet to warm up to
ambient temperature (at least 1 h) before connecting it to the mains power supply.
• The weight of the rolling cabinet is 40.0 kg (88.2 lb).
The weight of the rolling cabinet with transformer is 60.0 kg (132.3 lb).
• At least 2 people are required to unpack and position the rolling cabinet.
• When unpacking, grab the rolling cabinet from below. Do not use the drawer
handle as a carrying handle.
The adhesive film on the upper side of the rolling cabinet may not be removed
yet. First mount the foot holders (see Mounting the foot holders on p. 15).