Fault
Check
Cause
Solution
P600OE stops
during operation
Battery or mains
unit positioned
correctly in the
connection slot
and locked in
position
Battery or mains unit
not inserted properly
Remove the battery
or mains unit from the
pump and re-insert.
Mains unit temperature
too high as a result
of continuous use
and high ambient
temperature
Allow mains unit to cool
down (see operating
instructions of mains
unit)
Electric motor not
providing full power
Check connection
of mains unit
Mains unit not correctly
connected
Re-insert the mains unit
Cable losses in
extension cable or
cable drums too great
(electrical resistance)
Use a different suitable
extension cable or cable
reel or connect directly
to power supply
Mains unit fault
Use a battery or
different mains unit.
Check battery
charge status
Battery almost fl at
Charge battery
Battery defective
Replace battery
Fault in power supply to
the mains unit
Troubleshooting the
power supply
Use a different power
supply
P600OE defective
Shut down immediately
and have repaired by
authorised service
facility, authorised
dealer or directly by
LUKAS