Troubleshooting – Lukas P 600 OE User Manual
Page 26

26
11. Troubleshooting
Fault
Check
Cause
Solution
The PO600OE
switches
inadvertently to the
Eco mode when
working with a
rescue unit
The initiation sequence
period of 3 seconds
is not allowed to be
carried out
Switch the pump off
and on again. Use the
rescue unit only after
the initiation sequence
has been fully carried
out. Flashing light
in the on/off switch
goes to permanent
light (see also
Paragraph 7.1)
Electric motor does
not start following
activation of the
main switch
Battery or mains
unit positioned
correctly in the
connection slot
and locked in
position
Battery or mains unit
not inserted properly
Remove the battery
or mains unit from the
pump and re-insert.
Check connection
of mains unit
Mains unit cable not
connected
Connect mains unit
cable to the power
supply
Correct battery
fi tted?
Incorrect battery
Use stipulated battery
Battery fl at
Charge battery
Battery or mains unit
defective
Use a different battery
or mains unit.
Fault in power supply to
the mains unit
Troubleshooting the
power supply
Use a different power
supply
P600OE defective
Shut down immediately
and have repaired by
authorised service
facility, authorised
dealer or directly by
LUKAS