Important: Always perform calibration procedure after first turning the unit on and allowing it to warm
up and stabilize for 20 seconds. If possible please calibrate the scale on your SCALEPAD (included in the
box), mousepad or similar surface. Calibrate your scale at regular intervals if the scale is subject to
weather or altitude changes or if the display shows Out2. Incorrect calibration can occur if you do not
follow the steps exactly. If your scale does not perform accurately, please also try replacing your
batteries.
Correct Calibration weight: F2-125: 100g F2-400: 200g
NOTE: if you do not have access to a calibration weight you can purchase one at your local store or in
emergency situations you can use coins or weights (ie 20 US Nickels = 100g) as a calibration weight.
Calibration Procedure:
Place the scale on a Flat, very stable surface.
1) Turn the scale ON.
2) Press and hold the UNIT key for 3~5 seconds until the display
show CAL.
3) Press the UNIT key again, the display will flash CAL and
then 100.00 (200.0 on the F2-400)
4) Place a 100g (200g for the F2-400) weight on the tray, the display will show PASS and then 100.00
(200.0). Calibration is complete.
NOTE: if after calibration your scale does not read accurately, this indicates calibration error and the
calibration process should be repeated more slowly. Please calibrate on your scalepad, mousepad or
similar surface if possible.
FEATURES____________________________________________
* Power Up Segment Test: When first turning the unit on, all segments of the display will appear as
shown below. This display will remain for approximately 3 seconds and then reset to zero. The display
may also change brightness during this time and show a blank screen for a moment before 0.0 appears