Send the result immediately in basic unit, Send the stable result in current unit, In basic unit – RADWAG PUE C/31H/Z Indicator User Manual
Page 58: Send the result immediately, Frame format, Syntax: si cr lf, Possible answers
- 58 -
Frame format:
1 2-3
4
5 6 7-15 16 17
18
19
20
21
S space stability space sign
mass
space unit CR
LF
Example:
S CR LF – computer command
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done,
mass value in basic unit is returned.
18.4.5. Send the result immediately in basic unit
Syntax: SI CR LF
Possible answers:
SI_I CR LF
- command comprehended but cannot be executed at the
moment
MASS FRAME
- mass value in basic unit is returned
Frame format:
1 2
3
4
5
6 7-15
16 17 18 19 20 21
S I space stability space sign mass space
unit
CR LF
Example:
S I CR LF – computer command
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass
value in basic unit is returned immediately.
18.4.6. Send the stable result in current unit
Syntax: SU CR LF
Possible answers: