beautypg.com

Minimal mass threshold – RADWAG WTB 2000 Precision Balance User Manual

Page 25

background image

Pr_n noStAb

-

immediate printout
(not accessible in verified scales)

Pr_n StAb

-

sending stable results

Pr_n rEPL

-

automatic operation

Pr_n CntA

-

continuous transmission in basic unit

Pr_n Cntb

-

continuous transmission in present unit


Return to weighing:

see 11.2.2.



14.2. Minimal mass threshold

This function is necessary while working with automatic tare or automatic
operation or weighing animals
.

Automatic tarring will not be applied until the indication (gross) is lower
than the value inscribed in S_Lo parameter.

In automatic operation measurements (net) are sent via RS232 when the
indication is equal or greater than the value inscribed in S_Lo parameter.

Weighings animals is performer when the indication is equal or greater
than the value inscribed in S_Lo parameter.


Procedure:

• Enter the submenu and then:

- 25 -