Syringe diameters and rate limits – Next Advance SP300 User Manual
Page 14

NEXT>>>ADVANCE
www.nextadvance.com
SP300 Programmable Syringe Pump
Publication #1200-01
11
03/23/10
9.5
Syringe Diameters and Rate Limits
Next Advance
www.nextadvance.com
Model: SP300 syringe pump
Maximum and minimum flow rates
Syringe
Manufacturer
Syringe
(cc)
Inside
Diameter
(mm)
Maximum
Rate
(mL/hr)
Minimum
Rate
(µl/hr)
Maximum
Rate
(mL/min)
B-D
1
4.699
39.27
0.73
0.6545
3
8.585
131
2.434
2.184
5
11.99
255.6
4.748
4.261
10
14.43
370.3
6.876
6.172
20
19.05
645.4
11.99
10.75
30
21.59
829
15.4
13.81
60
26.59
1257
23.35
20.95
HSW
1
4.69
39.11
0.727
0.6519
Norm-Ject
3
9.65
165.6
3.076
2.76
5
12.45
275.6
5.119
4.594
10
15.9
449.6
8.349
7.493
20
20.05
714.9
13.28
11.91
30
22.9
932.6
17.32
15.54
50
29.2
1516
28.16
25.27
Monoject
1
5.74
58.59
1.088
0.9766
3
8.941
142.1
2.64
2.369
6
12.7
286.8
5.326
4.78
12
15.72
439.4
8.161
7.324
20
20.12
719.9
13.37
11.99
35
23.52
983.8
18.27
16.39
60
26.64
1262
23.44
21.03
140
38.00
47.69
42.80
Terumo
1
4.7
39.28
0.73
0.6547
3
8.95
142.4
2.646
2.374
5
13
300.5
5.581
5.009
10
15.8
443.9
8.244
7.399
20
20.15
722.1
13.41
12.03
30
23.1
949
17.63
15.81
60
29.2
1516
28.16
25.27
Air-Tite
10
15.9
449.6
8.349
7.493
20
20.25
729.2
13.55
12.15
30
22.5
900.3
16.72
15
50
29
1495
27.78
24.92
Syringe
(µL)
Inside
Diameter
(mm)
Maximum
Rate
(µl/hr)
Minimum
Rate
(µl/hr)
Hamilton
0.5
0.103
18.86
0.001
Microliter
1
0.146
37.91
0.001
2
0.206
75.47
0.002
5
0.326
189
0.004
SGE
0.5
0.1
17.78
0.001
1
0.15
40.01
0.001