Setup Configuration: Set Power Failure Mode
To change or view the setup configuration, the Pump must be stopped. Press and hold the
‘Diameter’/‘Setup’ key until the parameter, ‘PF’ is displayed. After 2 seconds, or when any non-arrow
key is pressed, the next parameter will be displayed . (See sec. 5.10, ‘Setup’ Key). Pressing an arrow key
under a value will change the parameter. The following will be displayed:
n
Power Failure Mode. Where ‘n’ is the current setting.: ‘0’ = Disabled, ‘1’ = Enabled.
When enabled, if the Pump was pumping when power to the pump was disrupted, the Pump will
automatically start pumping when power is reconnected to the pump. The accumulated volume will be
reset to 0. Pressing any key on the keypad while powering up the pump will stop the Pump from starting
in Power Failure Mode.
n
Low Noise Mode, Where ‘n’ is the current setting.: ‘0’ = Disabled, ‘1’ = Enabled.
A side affect of the SP300’s high precision micro-stepped motor driver is a high frequency resonance
sound at very low pumping speeds. This mode minimizes this sound by reducing the micro-stepping,
increasing pulsations.
8.
Rear of Pump
NEW ERA PUMP SYSTEMS, Inc.
WWW.SYRINGEPUMP.COM
(631) 249-1392
Farmingdale, NY [email protected]
NE-300
Figure 2: Rear of Pump
9.
Appendix
9.1
Accessories
9.1.1 Syringe Heater
Part#: SYRINGE-HEATER, see www.SyringeHeater.com for details.
Flexible heating pad that wraps around the syringe. Digital PID controller will heat a syringe to a set
temperature up to 100 C.
9.1.2 Firmware and Hardware Upgrade
Contact your dealer about upgrading your pump with other features from the NE-1000 Programmable
Syringe Pump family. See www.SyringePump.com