beautypg.com

Operation, Special power-up functions, Error and status messages – Next Advance SP300 User Manual

Page 10: Syringe inside diameter

background image

NEXT>>>ADVANCE

www.nextadvance.com

SP300 Programmable Syringe Pump

Publication #1200-01

7

03/23/10

diameter. Any new parameter value will take affect immediately upon being stored.

5.11

Special Power-Up Functions

The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.

5.11.1 Firmware Version Display

To display the pump’s firmware version, press the left-most up arrow key (

/

) while turning on

power to the pump. The display will show:

, or similar. Pressing any key will clear the

display.

5.11.2 Reset Pumping Parameters

To clear out the current Pump setups, press the right-most up arrow key (

) while turning on power to

the pump. The display will show

. Pressing any key will clear the display.

With pumps having as many complex features as the NE-1000 family, it is easy for a novice user
experimenting with the pump's setup to get the pump into a 'weird' state. Performing this reset function
will bring the pump out of a 'weird' state.

5.12

Error and Status Messages

Value entered is ‘Out Of Range’ of the pump’s operational limits. Check the pumping
rate and/or the syringe inside diameter setting is correct.

Key pressed is not currently applicable.

Indicates pumping rate or volume units change mode. The units LED's will also be
blinking. Press any up arrow key to change the units.

Indicates that the pump is purging. Displayed while holding down the 'Start/Stop' key.

Displayed while holding down the direction key and performing a reverse purge.

6.

Operation

Before the pump can be operated, the pumping data must be setup. The syringe inside
diameter and a non-zero pumping rate needs to be set. The operation of the pump can
then be started by pressing the ‘Start / Stop’ key.

6.1

Syringe Inside Diameter

The syringe inside diameter can only be set while the Pump is stopped. Use the up arrow keys to set the
diameter value. While the diameter value is being set, the ‘mm’ LED will blink. The new diameter value
is stored after pressing any non-arrow key, or after a 2 second delay.

Valid syringe diameters are from 0.1 mm to 50.0 mm. If the diameter is out of this range, the display will
show ‘oor’. Pressing any key restores the diameter display to its previous value. Changing the syringe
diameter will not zero the pumping rate. Section 9.5, “Syringe Diameters and Rate Limits”, is a
representative list, for reference, of syringe diameters for various syringe manufacturers and syringe sizes.

6.1.1 Volume Units: Default and How to Change

The units of the accumulated infusion volume is set according to the syringe diameter setting. If the
default volume units are changed (see next section), the selected volume units will remain in affect until a
reset function is performed.

From 0.1 to 14.0 mm

Syringes smaller than 10 ml:

Volume units are ‘

l’

From 14.01 to 50.0 mm

Syringes greater than or equal to 10 ml:

Volume units are ‘ml’