Troubleshooting and maintenance, Specifications mechanical & electrical, Operational – Next Advance SP300 User Manual
Page 13: Custom applications
![background image](/manuals/466044/13/background.png)
NEXT>>>ADVANCE
www.nextadvance.com
SP300 Programmable Syringe Pump
Publication #1200-01
10
03/23/10
9.2
Troubleshooting and Maintenance
Maintenance: Periodically, a small amount of all-purpose oil should be applied to the guide rods.
The mechanism should be kept clean to prevent impeded operation.
No other special maintenance or calibrations are needed
Pusher block makes a snap or click sound when the pump is started: This is a normal condition.
When the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive
screw. The sound heard is the drive-nut engaging on the drive screw.
9.3
Specifications
Mechanical & Electrical
Syringe sizes:
Up to 140 cc
Number of syringes:
1
Motor type:
Step motor
Motor steps per revolution:
400
Microstepping:
1/8 to 1/2 depending on motor speed
Advance per step:
0.2126 uM to 0.8504 uM depending on motor speed
Motor to drive screw ratio:
15/28
Drive screw pitch:
20 revolutions/”
DC connector:
2.1 mm, center positive
Voltage at DC connector:
12V DC at full load
Amperage:
750 mA at full load
Power supply type:
Unregulated linear external wall adapter, country and power
source specific
Power supply output rating:
12V DC @ 800 mA
Dimensions:
8 3/4” x 5 3/4” x 4 1/2” High
(22.86 cm x 14.605 cm x 11.43 cm)
Weight:
3.6 lbs. (1.63 kg)
Operational
Maximum speed:
3.7742 cm/min
Minimum speed:
0.004205 cm/hr
Maximum pumping rate:
1257 ml/hr with a B-D 60 cc syringe
Minimum pumping rate:
0.73
l/hr with a B-D 1 cc syringe
Maximum force:
35 lbs. at minimum speed, 18 lbs. at maximum speed
Syringe inside diameter range:
0.100 to 50.00 mm
9.4
Custom Applications
For specialized and OEM applications, contact your dealer
or
Next Advance
Inc. Custom modifications can be made to
the mechanics or the firmware.