46203-90
MINIMUM / MAXIMUM
The Temp Tracker stores minimum and maximum values for both
displays. To view minimum and maximum values, depress the
SET/ RESET key momentarily. Each time the key is depressed, the
display switches modes; first to
MIN then MAX then back to Cur-
rent values. To reset all values, press and hold the SET / RESET
key until the unit beeps.
The Temp Tracker may be set to automatically cycle through
MIN,
MAX and Current values. This feature is selected during SETUP.
Minimum and maximum values reset if power is interrupted.
REMOTE DISPLAYS
The Temp Tracker may be used as a remote or slave display by
selecting InP SEr during SETUP. Remote displays are connected
to the Main display using the RS-485 terminals. Connect like
terminals as shown in the wiring diagram. Remote displays show
exactly the same information as the Main display. MIN / MAX and
SETUP functions are controlled at the Main unit only. Brightness is
adjusted at each display independently.
VOLTAGE OUTPUTS
The Temp Tracker offers calibrated voltage outputs for both chan-
nels. This feature allows the use of recorders and other devices.
Full scale voltage for each channel is 5.00 VDC.
ERROR MESSAGES
The Temp Tracker will detect and indicate certain errors. Once cor-
rected, the error indication disappears.
Display
Error
LEFT RIGHT
- - -
- - -
4-20 mA signal from sensor is not present or is
outside acceptable range.
SEr
Err
Unit is set up to receive serial input signal, but
no data is being received. Verify that main unit
is
operating. Check cable connections.
WARRANTY
The Temp Tracker is warranted to be free of defects in materials and
workmanship for a period of 12 months from the date of purchase.
This warranty is limited to repair or replacement of defective unit.
SPECIFICATIONS
• Size
144 x 144 x 36 mm (5.65 x 5.65 x 1.4 in)
• Panel Cutout
138 x 138 mm (5.43 x 5.43 in)
• Sensor Type
YOUNG Model 41382LC2 RH / T
YOUNG Model 41342LC T only
• Input Signal
T
4-20mA = -50 to +50°C
RH
4-20mA = 0 to 100%
• Serial I/O
RS-485, binary data block,
16 blocks per second
• Accuracy
Temperature ±0.1 °C (0.2°F)
Relative Humidity ±0.1 %
Dew Point ±0.5 °C (0.9 °F)
(These accuracies are for the Temp Tracker
only and are independent of sensor accuracy.)
• Voltage Output 0.00 to 5.00 VDC
Scale
Temperature and Dew Point:
-50 to 50 °C, -50 to 150 °F
Relative Humidity: 0 to 100%
• Alarm relays
Normally open, non-latching,
5A resistive, 2A inductive @ 24 VAC or
30 VDC
• Power
requirement
12 to 30 VDC, 4.5 watts
• Weight
0.75 lbs (0.34 kg)
• Accessories
Model 06280 Rack Mounting Panel
CE COMPLIANCE
This product has been tested and shown to comply with European CE
requirements for the EMC Directive (see Declaration of Conformity
below). Shielded cable must be used.
Declaration of Conformity
R. M. Young Company
2801 Aero Park Drive
Traverse City, MI 49686 USA
Model 46203 TEMP TRACKER
The undersigned hereby declares on behalf of R. M.
Young Company that the above-referenced product, to
which this declaration relates, is in conformity with the
provisions of:
Council Directive 2004/108/EC (December 15, 2004)
on Electromagnetic Compatibility
David Poinsett
R&D Manager