INTRODUCTION
The Multi-Plate Radiation Shield protects temperature and
relative humidity sensors from error-producing solar radiation
and precipitation. This shield relies on a combination of plate
geometry, material and natural ventilation to provide effective
shielding.
INSTALLATION
The Multi-Plate Radiation Shield is designed to be mounted on
a relatively flat open area. For best performance, the shield
should be placed in a location with the following characteristics:
- Good air circulation around shield.
- Away from large masses (asphalt, masts, solar panels)
especially metal items.
- Away from building exhaust vents, electrical machinery
and motors.
- Away from standing water, water fountains and
sprinklers.
MAINTENANCE
The Multi-Plate Radiation Shield requires no regular maintenance
in normal use. Dirt or dust accumulation on the plates and bracket
can be cleaned with soap and water.
Do not use solvents.
SPECIFICATIONS
Sensor Mounting:
Model 41003:
Universal adapter for sensors up to 16mm diameter.
Model 41003P-24:
A 24mm adapter for standard YOUNG sensors.
Model 41003P:
Custom sensor adapter that can be machined for
sensors up to 26mm diameter.
Radiation Error: 1080 W/m
2
intensity
0.4°C (0.7°F) RMS @ 3 m/s (6.7 mph)
0.7°C (1.3°F) RMS @ 2 m/s (4.5 mph)
1.5°C (2.7°F) RMS @ 1 m/s (2.2 mph)
Construction:
UV stabilized white thermoplastic plates.
Stainless Steel Plate Studs.
Aluminum mounting bracket, white powder coated.
Stainless steel U-bolt mounting clamp.
Dimensions:
13 cm (5.1 in) D x 27 cm (10.6 in) H
Mounting fits vertical pipe 25-50 mm (1-2 in) diameter
Weight:
Net wt 0.7 kg (1.6 lb)
Ship wt 1.4 kg (3.2 lb)
MODEL 41003
MULTI-PLATE RADIATION SHIELD
INSTRUCTION SHEET 41003-90
REV C012114