Young Sensors Relative Humidity & Temperature Probes Models 41382LC2 User Manual
Young Sensors
MODEL 41382LC2
RELATIVE HUMIDITY
/
TEMPERATURE PROBE
with CURRENT OUTPUTS
INSTRUCTION SHEET 41382LC2-90
REV B102613
R. M. YOUNG COMPANY
2801 Aero Park Drive , Traverse City, Michigan 49686 USA
TEL (231) 946-3980 FAX (231) 946-4772
I
NTRODUCTION
The Model 41382LC2 Relative Humidity/Temperature Probe
combines high accuracy humidity and temperature sensors in a
single probe. The output signal is 4-20 mA. RH range is 0-100%.
Temperature range is -50 to +50°C.
INSTALLATION
Install the probe in a naturally ventilated or fan aspirated radiation
shield. Mount probe and shield in a location with good air
circulation clear of large thermal masses (buildings, pavement,
solar panels, etc.), exhaust vents, electrical machinery and motors,
water fountains and sprinklers. Typical wiring diagrams appear on
next page.
MAINTENANCE
The probe provides years of service with minimal maintenance.
Humidity calibration, which may drift slightly with time, can be
checked or restored at the factory.
Periodically clean the humidity sensor filter when used in areas of
high dust or contamination (smokestacks, seawater, etc.) Soak in
clean water or use a mild soap solution. DO NOT USE SOLVENTS.
CE COMPLIANCE
This product complies with the European CE directive on EMC
compatibility. Shielded cable must be used.
WARRANTY
This product is warranted to be free of defects in materials
and construction for a period of 12 months from date of
initial purchase. Liability is limited to repair or replacement of
defective item. A copy of the warranty policy may be obtained
from R. M. Young Company.
SPECIFICATIONS
RELATIVE HUMIDITY:
Measuring range:
0-100% RH
Accuracy at 23°C:
±1%
Stability:
Better than ±1%RH per year
Response Time:
10 seconds (Without Filter)
Sensor type:
Rotronic Hygromer
TEMPERATURE:
Measuring Range:
-50 to +50°C
Accuracy at 23°C:
±0.3°C
Response Time:
10 seconds (Without Filter)
Sensor type:
Platinum RTD
W
Output signal:
4-20 mA
Power Required:
5-30 VDC, 46 mA max
Recommended Cable:
2 Pair shielded 22AWG, Young 18723
Recommended Shields:
Young Model 43502 Aspirated Radiation Shield
Young Model 41003P Multi-Plate Radiation Shield
Declaration of Conformity
R. M. Young Company
2801 Aero Park Drive
Traverse City, MI 49686 USA
Model 41382LC2 RH/Temp Probe
The undersigned hereby declares on behalf of R. M. Young
Company that the above-referenced product, to which this
declaration relates, is in conformity with the provisions of:
Council Directive 2004/108/EC (December 15, 2004)
on Electromagnetic Compatibility
David Poinsett
R&D Manager
7.4in
188mm
5.4in
136mm
.94in
24mm
2.5in
62mm
3.4in
86.5mm