beautypg.com

Warranty, Ce compliance – Young Heavy Duty Wind Monitor-HD-Alpine Model 05108-45 User Manual

Page 5

background image

Page 4

05108-45-90(A)

WARRANTY

This product is warranted to be free of defects in materials and

construction for a period of 12 months from date of initial purchase.

Liability is limited to repair or replacement of defective item. A copy

of the warranty policy may be obtained from R. M. Young Company.

CE COMPLIANCE

This product has been tested and shown to comply with European

CE requirements for the EMC Directive. Please note that shielded

cable must be used.

VERTICAL SHAFT BEARING REPLACEMENT:

Vertical shaft bearings are much larger than the anemometer

bearings. Ordinarily, these bearings will require replacement less

frequently than anemometer bearings. Check bearing condition

using a Model 18331 Vane Torque Gauge. Without it, a rough

check can be performed by holding the instrument with the vane

horizontal and placing a 3 gm weight near the aft edge of the fin.

A U.S. penny weighs about 3 gm and is convenient for this check.

Failure to rotate downward indicates the vertical bearings need

replacement. Repeat this test at different positions to check full

bearing rotation.

Since this procedure is similar to POTENTIOMETER

REPLACEMENT, only the major steps are listed here.

1. REMOVE MAIN HOUSING

2. UNSOLDER TRANSDUCER WIRES AND REMOVE

TRANSDUCER ASSEMBLY

Loosen set screws at base of transducer assembly and

remove entire assembly from vertical shaft.

3. REMOVE VERTICAL SHAFT BEARING ROTOR by sliding it

upward off vertical shaft.

4. REMOVE OLD VERTICAL BEARINGS AND INSTALL NEW

BEARINGS. When inserting new bearings, be careful not to

apply pressure to bearing shields.

5. REPLACE VERTICAL SHAFT BEARING ROTOR.

6. REPLACE TRANSDUCER & RECONNECT WIRES

7. REPLACE MAIN HOUSING

8. ALIGN VANE

9. REPLACE NOSE CONE

Declaration of Conformity

R. M. Young Company

2801 Aero Park Drive

Traverse City, MI 49686 USA

Model 05108-45 Wind Monitor - HD Alpine

The undersigned hereby declares on behalf of R. M.

Young Company that the above-referenced product, to

which this declaration relates, is in conformity with the

provisions of:

Council Directive 2004/108/EC (December 15, 2004)

on Electromagnetic Compatibility

David Poinsett

R&D Manager