Warranty, Ce compliance – Young Serial Output Wind Monitor Model 09101 User Manual
Page 6

Page 5
09101-90(G)
Declaration of Conformity
R. M. Young Company
2801 Aero Park Drive
Traverse City, MI 49686 USA
Model 09101 Wind Monitor-SE
The undersigned hereby declares on behalf of R. M.
Young Company that the above-referenced product, to
which this declaration relates, is in conformity with the
provisions of:
Council Directive 2004/108/EC (December 15, 2004)
on Electromagnetic Compatibility
David Poinsett
R&D Manager
WARRANTY
This product is warranted to be free of defects in materials and construction
for a period of 12 months from date of initial purchase. Liability is limited
to repair or replacement of defective item. A copy of the warranty policy
may be obtained from R. M. Young Company.
CE COMPLIANCE
This product has been tested and shown to comply with European CE
requirements for the EMC Directive (see Declaration of Conformity
below). Please note that shielded cable must be used.
9. REPLACE MAIN HOUSING
a) Position main housing over bearing rotor. Be careful to align
indexing key inside main housing with slot in rotor.
b) Turn direction-adjust thumbwheel until notch in coupling is away
from front opening and aligned with ridge inside housing. Set
screw in coupling should face the front opening.
c) With coupling oriented, continue to push main housing onto
bearing rotor until latch locks into place with a click.
10. ALIGN VANE
a) Connect sensor to indicator.
b) Install sensor on vane angle fixture (Young Model 18112 or
equivalent) with junction box at 180° or South position.
c) Align sensor to known angular position. If indicator output varies
more than ±1 ° from known angle, loosen setscrew in direction
adjust thumbwheel and slowly turn thumbwheel until correct
output value is obtained. Tighten setscrew.
d) Verify correct angular values at other vane positions.
11. REPLACE NOSE CONE
a) Screw nose cone into main housing until o-ring seal is seated. Be
careful to avoid cross-threading.