6 Commissioning, operation
40 WIKA Operating Instruction, Pascal 100 and Pascal 100/IS
GB
The next screen shows the Span parameter (measuring range). For default the sensor has its
maximum range equal to the measuring range (Span). If the Span has to be redefined, or to be
reduced, the same can be changed by using a numeric keyboard. The redefinition of the Span is
used during a calibration for comparison where a relationship between a REF and a DUT channel
has to be done.
Figure 32
– Sensor range setup
Now it is possible to finalize the procedure by pressing the YES button or come back to the
previous step pressing NO.
Figure 33 - Channel setup confirmation
Pressing the key YES, the channel is set up with the selected parameters and the screen with four
channels is displayed again.
Figure 34
– Main display with pressure measurement
- 890.09.2190 (44 pages)
- A-10 (96 pages)
- A2G-50 (52 pages)
- A2G-55 (36 pages)
- AC-1 (88 pages)
- C-2 (24 pages)
- D-10-7 (112 pages)
- D-20-9 (35 pages)
- D-20-9 (51 pages)
- DG-10 (112 pages)
- DP-10 (44 pages)
- DPT-10 (92 pages)
- DPT-10 (96 pages)
- GCS-1 (76 pages)
- HP-2 (84 pages)
- IL-10 (31 pages)
- IPT-10 (36 pages)
- IPT-10 (56 pages)
- IPT-10 (48 pages)
- IPT-10 (28 pages)
- IS-20-F (7 pages)
- IS-20-F (43 pages)
- IS-20-H (7 pages)
- UT-10 (42 pages)
- UT-10 (52 pages)
- IUT-10 (78 pages)
- LH-10 (60 pages)
- LH-20 (60 pages)
- LS-10 (60 pages)
- MG-1 (92 pages)
- MH-1 (11 pages)
- MH-2 (9 pages)
- MHC-1 (84 pages)
- N-10 (35 pages)
- O-10 (108 pages)
- OT-1 (60 pages)
- P-30 (92 pages)
- PSA-31 (124 pages)
- PSD-30 (128 pages)
- R-1 (92 pages)
- S-10 (35 pages)
- S-11 (102 pages)
- S-20 (52 pages)
- S-20 (96 pages)
- SL-1 (21 pages)