Pressure module – WIKA Pascal 100 User Manual
Page 19
6 Commissioning, operation
WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 19
GB
6.1.2.3 Pressure module
The pressure module is installed only at factory location in accordance with the Pascal 100 or
Pascal 100/IS configuration required. It is possible to install two pressure modules, each one can
support up to two internal and one external transducers. In total Pascal 100 can accommodate four
internal and two external transducers.
Transducer’s selection with range and their resolutions are
in the following table. If requested, one of the standard pressure modules can be changed with one
high accuracy pressure module (with one high accuracy transducer).
Internal transducers *
Range
Precision (% FS)
Accuracy (% FS)
Typical Resolution
Gauge
-
60…+60 mbar
0,08
0,1
0,01 mbar
-
500…+500 mbar
0,015
0,025
0,01 mbar
-
900…+1500 bar
0,015
0,025
0;01 mbar
0…7 bar
0,015
0,025
0,1 mbar
0…21 bar
0,015
0,025
0;1 mbar
0…50 bar
0,015
0,025
1 mbar
0…100 bar
0,015
0,025
1 mbar
Absolute
600…1300 mbar abs.
0,015
0,025
0,01 mbar
0…1500 mbar abs.
0,015
0,025
0,01 mbar
0…2500 mbar abs
0,015
0,025
0,01 mbar
0…2500 mbar abs
0,010
0,015
0,01 mbar
0…5 bar abs.
0,015
0,025
0,1 mbar
0…7 bar abs.
0,015
0,025
0,1 mbar
0…21 bar abs.
0,015
0,025
0,1 mbar
0…81 bar abs.
0,015
0,025
1 mbar
0…100 bar abs.
0,015
0,025
1 mbar
External transducers *
Range
Precision (% FS)
Accuracy (% FS)
Typical Resolution
Gauge
-
60…+60 mbar
0,1
0,15
0,01 mbar
-
500…+500 mbar
0,015
0,025
0,01 mbar
-
900…+1500 bar
0,015
0,025
0;01 mbar
0…7 bar
0,015
0,025
0,1 mbar
0…21 bar
0,015
0,025
0;1 mbar
0…50 bar
0,015
0,025
1 mbar
0…100 bar
0,015
0,025
1 mbar
0…200 bar
0,015
0,025
10 mbar
0…400 bar
0,015
0,025
100 mbar
0…700 bar
0,025
0,05
100 mbar
0…1000 bar
0,025
0,05
100 mbar
Absolute
0…1500 mbar abs.
0,015
0,025
0,01 mbar
0…2500 mbar abs
0,015
0,025
0,01 mbar
0…5 bar abs.
0,015
0,025
0,1 mbar
0…7 bar abs.
0,015
0,025
0,1 mbar
0…21 bar abs.
0,015
0,025
0,1 mbar
0…81 bar abs.
0,015
0,025
1 mbar
0…100 bar abs.
0,015
0,025
1 mbar
* Other ranges available on request