beautypg.com

Design and function, 2 dimensions in mm, Flow-through housing – WIKA TW61 User Manual

Page 8

background image

11593963.03 03/2014 GB/D/F/E

8

WIKA operating instructions thermowell, model TW61

GB

Pt100

Output signal

4 ... 20 mA

1) In the event of replacement, select a resistance thermometer

measuring insert model TR22-B.

The corresponding sensor length, A(l

5

), is calculated as follows:

Insertion length U

1

+ neck tube length M + 5 mm pre-load +

10 mm head dimension

Output signal

4.2 Dimensions in mm

Flow-through housing

4. Design and function

11610565.01

1)

1) In the event of replacement,

calculate sensor length, A,

as follows:

A

TR21-B

= U

1

+ M + 3 mm

11528266.01

14006093.02

For model TR21-B miniature resistance thermometer

For model TR22-B resistance thermometer

pr

e-load

1)