beautypg.com

Gb 8. technical data, Approval german lloyd (optional) – WIKA 891.34.2189 User Manual

Page 10

background image

GB

8. Technical data

Additional or deviating technical data

Pressure ranges

bar 0 ... 0.25 to 0 ... 10

Output signal

4 ... 20 mA, 2-wire or 0 ... 20 mA, 3-wire current limit I < 32 mA

Permissible ambient
temperature

°C

-10 … +70

EMC (electro-magnetic
compatibility)

Interference emission per

EN 50 081-1 (March 93) and
EN 50 081-2 (March 94),

interference immunity per

EN 50 082-2 (March 95)

ESD

kV

± 8

contact discharge

IEC 1000-4-2

Electromagnetic fields

V/m 10

80 % AM, 1 kHz, 0.01 ... 1000 MHz

IEC 1000-4-3

Burst

kV

± 2

coupling clamp

IEC 1000-4-4

Conducted HF-disturbance V

3

80 % AM, 1 kHz, 0.01 ... 100 MHz

IEC 1000-4-6

Surge

kV

± 0.5 symmetrically

IEC 1000-4-5

kV

± 1

asymmetrically, R i = 42 Ohm

kV

± 1

symmetrically

kV

± 2

asymmetrically, R i = 42 Ohm, with surge
protection only e.g. model MM-DS/x-
NFE(L), firm Dehn & Söhne or equivalent

Conducted NF-disturbance Veff

3

0.05 ... 10 kHz

IEC 945

Vibration test Fc
2 ... 25 Hz, +/- 1.6 mm

%

< 2.5 error

IEC 68-2-6

25 ... 100 Hz, 4 g

%

< 2.5 error

Approval German Lloyd (optional)

2279453 08/2009 GB/D/F

WIKA Operating Instructions Differential Pressure Transmitter DELTA-trans

10