beautypg.com

WIKA IPT-11 User Manual

Page 27

background image

Proceed as follows:

1

Edit the mbar value in the menu item "zero" with [OK].

Zero

000

.0 %

P

=

+0000

.0 mbar

0000.0 mbar

2

Set the requested mbar value with [+] and [->].

3

Confirm with [+] and move to span adjustment with [->].

The zero adjustment is finished.

Information:
The zero adjustment shifts the value of the span adjustment. The span,
i.e. the difference between these values, however, remains unchan-
ged.

Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.

If the adjustment ranges are exceeded, the message "Outside
parameter limits
" appears. The editing procedure can be aborted with
[ESC]

or the displayed limit value can be accepted with [OK].

Proceed as follows:

1

Edit the mbar value in the menu item "span" with [OK].

Span

100

.0 %

P

=

+1000

.0 mbar

0000.0 mbar

Information:
The displayed pressure for 100 % corresponds to the nominal
measuring range of the sensor (in the above example 1 bar =
1000 mbar).

2

Set the requested mbar value with [->] and [OK].

3

Confirm with [OK] and move to the menu overview with [ESC].

The span adjustment is finished.

Information:
For an adjustment with pressure, simply enter the actual measured
value indicated at the bottom of the display.

Carrying out zero ad-
justment

Carrying out span ad-
justment

Process pressure transmitter IPT-1* Vers. 3.0 - metalic sensor • 4 … 20 mA

27

6 Set up with the indicating and adjustment module

37300

-

EN

-100302