3-1-2 TM-2430 Series Error Codes

ERROR CODE

POSSIBLE CAUSE

CORRECTION

E00

Clock parameters have not been
set.

Enter clock parameters.

E03

An error code is displayed
without cuff inflation.

Exhaust the air from the cuff completely.

E04

Measurement is stopped due to a
low battery.

Replace with new batteries. Auto mode
is quit. Restart the Auto mode if you use
it.

E05

Inflation pressure does not reach
the target pressure.

Wrap the cuff and connect to the
monitor. If the error persists, there may
be an air leak and repair is necessary.

E06

Pressurization greater than 320
mmHg.

Do not move and try to relax during the
measurement. If you can not clear the
error, repair is necessary.

E07

Controlled

stop

using

STOP

key

Do not press the STOP key when
unnecessary.

E08, E10

Pulse can not be detected.

Do not move and try to relax during the
measurement.

E20

Pulse less than 35 beats or more
than 200 beats per minute.

Measure the blood pressure by other
methods.

E21

Diastolic greater than 160 mmHg
or less than 40 mmHg.

Measure the blood pressure by other
methods.

E22

Systolic greater than 280 mmHg
or less than 60 mmHg.

Measure the blood pressure by other
methods.

E23

DSD greater than 150 mmHg or
less than 10 mmHg.
(DSD = difference between
systolic blood pressure and
diastolic blood pressure)

Measure the blood pressure by other
methods.

E30, E31

Air leaks in the cuff assembly.

Repair is necessary.

E32

Clock error

If the error persists, repair is necessary.

E50

Pressure offset error

Repair is necessary.

E52

Memory error

Repair is necessary.

E53

Defective battery contact

Replace batteries correctly. If the error
persists, repair is necessary.

E55, E56, E57

Exhaust error

Relax and do not move during the
measurement. If this error occurs
frequently, repair is necessary.

E60

Interval setting error

Enter the interval parameters correctly.

E70, 71, 72, 73

RS-232C error

Re-connect the communication cable. If
the error persists, repair is necessary.

E74

Voltage decreases during
communication.

Replace batteries with new ones and
restart communication.

E75

Protocol error due to external
equipment

Re-connect the communication cable. If
the error persists, repair is necessary.