3-1-2 TM-2430 Series Error Codes
ERROR CODE
POSSIBLE CAUSE
CORRECTION
E00
Clock parameters have not been
set.
Enter clock parameters.
E03
An error code is displayed
without cuff inflation.
Exhaust the air from the cuff completely.
E04
Measurement is stopped due to a
low battery.
Replace with new batteries. Auto mode
is quit. Restart the Auto mode if you use
it.
E05
Inflation pressure does not reach
the target pressure.
Wrap the cuff and connect to the
monitor. If the error persists, there may
be an air leak and repair is necessary.
E06
Pressurization greater than 320
mmHg.
Do not move and try to relax during the
measurement. If you can not clear the
error, repair is necessary.
E07
Controlled
stop
using
STOP
key
Do not press the STOP key when
unnecessary.
E08, E10
Pulse can not be detected.
Do not move and try to relax during the
measurement.
E20
Pulse less than 35 beats or more
than 200 beats per minute.
Measure the blood pressure by other
methods.
E21
Diastolic greater than 160 mmHg
or less than 40 mmHg.
Measure the blood pressure by other
methods.
E22
Systolic greater than 280 mmHg
or less than 60 mmHg.
Measure the blood pressure by other
methods.
E23
DSD greater than 150 mmHg or
less than 10 mmHg.
(DSD = difference between
systolic blood pressure and
diastolic blood pressure)
Measure the blood pressure by other
methods.
E30, E31
Air leaks in the cuff assembly.
Repair is necessary.
E32
Clock error
If the error persists, repair is necessary.
E50
Pressure offset error
Repair is necessary.
E52
Memory error
Repair is necessary.
E53
Defective battery contact
Replace batteries correctly. If the error
persists, repair is necessary.
E55, E56, E57
Exhaust error
Relax and do not move during the
measurement. If this error occurs
frequently, repair is necessary.
E60
Interval setting error
Enter the interval parameters correctly.
E70, 71, 72, 73
RS-232C error
Re-connect the communication cable. If
the error persists, repair is necessary.
E74
Voltage decreases during
communication.
Replace batteries with new ones and
restart communication.
E75
Protocol error due to external
equipment
Re-connect the communication cable. If
the error persists, repair is necessary.