beautypg.com

Bp data list, 6-2 bp data list – A&D TM-2430-13 User Manual

Page 20

background image

19

2-6-2 BP Data List

The BP Data List function displays individual blood pressure measurement within an open data file in

tabular form. With the BP Data List, you may input measurement comments, exclude/include a

measurement within the graphics and statistics or link a measurement to the beginning of the Trends

Screen.

The following table defines each heading within the Blood Pressure Data List:

Item

Description

#

Measurement number

DATE/TIME

Date and time of the measurement

SYS

Systolic blood pressure

DIA

Diastolic blood pressure

MAP

Mean arterial blood pressure, obtained by MAP=DIA+(SYS-DIA)/3

PUL

Pulse

DP

Double Product, obtained by DP=SYS

PUL/100

STATUS

The status line has 3 positions. When filled with a character other than “–”,
each denotes a specific occurrence or condition.

Position 1
BP method,
O or K

Denotes the BP method used for the specific measurement. “K” refers to
Korotkoff method, “O” refers to oscillometric method.

Position 2
E (Event)

Indicates that data was measured when the monitor’s EVENT button was
activated.

Position 3
Error code

Denotes the specific error code that occurred during the monitoring session,
see “3-1 Measurement Errors”.

Exclude

When this column has an “X”, the measurement has been excluded from the
statistics and graphs.

Comments

Space provided for user to enter patient diagnostic notations up to 20
characters.

Vertical scroll

bar

The vertical scroll bar on the right hand side of the screen helps scroll through
the measurements. This may also be performed with the up and down arrow
keys, as well as the mouse.

Save button

Saves any edits, i.e. comments, exclusions or inclusions.

Close button

Closes the screen.
Note: Edits (comments, exclusions or inclusions) will not be saved.

Help button

Opens the Blood Pressure Data List help screen.