65

same reference is used for both the pH and the ORP sensors. Both

pH and ORP will indicate 0 for a neutral solution. Calibration at zero

compensates for error in the reference junction.

A zero calibration solution for ORP is not practical, so the Ultrameter III

uses the offset value determined during calibration to 7 in pH calibration

(pH 7 = 0 mV). Sensitivity of the ORP surface is fixed, so there is no gain

adjustment either.

5. Sources of Error

The basics are presented in pH and ORP/Free Chlorine, pg. 62, because

sources of error are much the same as for pH. The junction side is the

same, and though the platinum surface will not break like the glass pH

surface, its protective glass sleeve can be broken. A surface film will

slow the response time and diminish sensitivity. It can be cleaned off

with detergent or acid, as with the pH glass.

C. Free Chlorine

1. Free Chlorine as an Indicator

Chlorine is a germicide used in the disinfection of drinking water,

wastewater, and recreational waters. It is dissolved in water and reacts

with water to form the hypochlorite ion and hypochlorous acid, the

concentrations of which are directly dependent on pH.

Some chlorine is initially consumed by organic matter and metal ions

present in solution. The amount of chlorine a particular solution consumes

is called chlorine demand. The amount of chlorine that remains beyond

this initial consumption is called free chlorine. Enough chlorine must be

added to solution so that residual levels of free chlorine have ongoing

germicidal properties. The 9P is designed to measure these residuals.

The 9P does NOT measure combined chlorine (compounds of chlorine

and nitrates that have no germicidal benefits), total chlorine (the sum of

free and combined chlorine), and chloramines (compounds of chlorine

and ammonia).

2. Free Chlorine Units

The 9P converts Oxidation Reduction Potential measured in mV to ppm

free chlorine using a proprietary algorithm developed by the Myron L

Company. This algorithm is based on a conversion curve that assumes

a pH of 5-9 in water sanitized by chlorine only.

The Myron L Company has been able to increase the accuracy of low-

range ppm free chlorine readings through extensive bench testing.

This feature allows the 9P to measure a dynamic range of sanitizer

concentrations that is wider than the range of a colorimetric test kit.