beautypg.com

Myron L 9PTK User Manual

Page 32

background image

28

e.

Press

once to confirm new value and end the calibration

sequence for this particular solution type. If another solution

type is also to be measured, change solution type now and

repeat this procedure.

2. User Calibration Conductivity/TDS

Instrument must be in User mode, see Solution Selection, pg. 22.
a.

Rinse conductivity cell three times with YOUR standard.

b.

Refill conductivity cell with same standard.

c.

Press

or

, then press twice in COND/three times

in TDS. The “cal” icon will appear on the display.

d.

Press or to step the displayed value toward the standard’s

value or hold a key down to scroll rapidly through the reading.

e.

Press once to confirm new value and end the calibration

sequence for this particular solution type.

3. Resistivity Calibration

Resistivity is the reciprocal of Conductivity. To calibrate resistivity,

calibrate conductivity for the solution type you wish to measure (ref.

Conductivity or TDS Calibration, pg. 27).

4. Reloading Factory Calibration (Cond or TDS)

If calibration is suspect or known to be incorrect, and no standard

solution is available, the calibration value can be replaced with the

original factory value for that solution. This “FAC” value is the same for

all Ultrameter IIIs, and returns you to a known state without solution

in the cell. The “FAC” internal electronics calibration (which bypasses

the electrodes and cell) is not intended to replace calibration with

conductivity/TDS standard solutions. If another solution type requires

resetting, change solution type and repeat this procedure.

a.

Press

or

.

b.

Press

. (If in “User” solution mode — press cal key

twice if in Conductivity, and three times if in TDS to skip over

tempco and ratio adjustments.)

c.

Press

key until “Fac” appears and release.