beautypg.com

Myron L 9PTK User Manual

Page 68

background image

64

5. Sources of Error

The basics are presented in pH and ORP/FREE CHLORINE, pg. 62.

a. Reference Junction

The most common sensor problem will be a clogged junction because

a sensor was allowed to dry out. The symptom is a drift in the “zero”

setting at 7 pH. This is why the Ultrameter III 9P does not allow more

than 1 pH unit of offset during calibration. At that point the junction is

unreliable.

b. Sensitivity Problems

Sensitivity is the receptiveness of the glass surface. A film on the surface

can diminish sensitivity and cause a long response time.

c. Temperature Compensation

pH sensor glass changes its sensitivity slightly with temperature, so the

further from pH 7 one is, the more effect will be seen. A pH of 11 at 40°C

would be off by 0.2 units. The Ultrameter III 9P senses the sensor well

temperature and compensates the reading.

B. ORP/Oxidation-Reduction Potential/REDOX

1. ORP as an Indicator

ORP is the measurement of the ratio of oxidizing activity to reducing

activity in a solution. It is the potential of a solution to give up electrons

(oxidize other things) or gain electrons (reduce).
Like acidity and alkalinity, the increase of one is at the expense of the

other, so a single voltage is called the Oxidation-Reduction Potential,

with a positive voltage showing, a solution wants to steal electrons

(oxidizing agent). For instance, chlorinated water will show a positive

ORP value.

2. ORP Units

ORP is measured in millivolts, with no correction for solution temperature.

Like pH, it is not a measurement of concentration directly, but of activity

level. In a solution of only one active component, ORP indicates

concentration. Also, as with pH, a very dilute solution will take time to

accumulate a readable charge.

3. ORP Sensors

An ORP sensor uses a small platinum surface to accumulate charge

without reacting chemically. That charge is measured relative to the

solution, so the solution “ground” voltage comes from a reference

junction - same as the pH sensor uses.

4. The Myron L ORP Sensor

Figure 34, pg. 63, shows the platinum button in a glass sleeve. The