Myron L 9PTK User Manual
Page 53
![background image](/manuals/460916/53/background.png)
49
software application.
2.
Verify that the port selected matches the COM port number
noted (first time only). This is the outgoing COM port on
Windows XP.
3.
In the U2CI application, click on the data download button.
A data transfer bar will appear while the data is being
downloaded.
Once downloaded, the data may be manipulated, printed or stored within
the Myron L U2CI application, or the data may be exported to another
more powerful spreadsheet, such as Excel*.
Please Note: Although the Myron L Company has performed extensive
testing, we cannot guarantee compatibility of all applications and formats.
We suggest testing your application and format for compatibility before
relying on it.
Additional features such as assigning a name to the unit, setting time
and date and erasing data are available. See U2CI software installation
CD or visit our website for the latest instructions:
http://myronl.com/main/U2CI_Application_DL.htm
4.
Upon completion, click on the “disconnect” icon.
5.
Turn off Ultrameter III PC download mode by selecting any
measurement function. Failure to do so will reduce battery life.
* Windows 2000, 2007, XP & Vista are registered trademarks of Microsoft Corporation.
** Macintosh OS9.2 & OSX are registered trademarks of Apple Computer Inc.
Xvii. care and maintenance
Ultrameter
•
III
s should be rinsed with clean water after use.
Solvents should be avoided.
•
Do not drop, throw or otherwise strike the instrument. Shock
•
damage may cause instrument failure.
A. Temperature Extremes
Solutions in excess of 71°C/160°F should not be placed in the cell
cup area; this may cause damage. The pH sensor may fracture if the
Ultrameter III temperature is allowed to go below 0°C/32°F. Care should
be exercised not to exceed rated operating temperature.
Leaving the Ultrameter III in a vehicle or storage shed on a hot day
can easily subject the instrument to over 66°C/150°F. This will void the
warranty.