Myron L 6P and 4P User Manual
Page 46
42
over this change of 10°C. The solution is then said to have a tempco of
2.2 %/°C.
Tempco always varies among solutions because it is dependent on their
individual ionization activity, temperature and concentration. This is why
the Ultrameter II features mathematically generated models for known
salt characteristics that also vary with concentration and temperature.
B. Finding the Tempco of an Unknown Solution
One may need to measure compensated conductivity of some solution
unlike any of the 3 standard salts. In order to enter a custom fixed
tempco for a limited measurement range, enter a specific value through
the “USER” function. The tempco can be determined by 2 different
methods:
1.
Heat or cool a sample of the solution to 25°C, and measure its
conductivity. Heat or cool the solution to a typical temperature
where it is normally measured. After selecting USER function,
set the tempco to 0 %/°C as in Disabling Temperature
Compensation, pg. 13 (No compensation). Measure the new
conductivity and the new temperature. Divide the % decrease
or increase by the 25°C value. Divide that difference by the
temperature difference.
2.
Heat or cool a sample of the solution to 25°C, and measure its
conductivity. Change the temperature to a typical measuring
temperature. Set the tempco to an expected value as in User
Programmable Temperature Compensation, pg. 12. See if the
compensated value is the same as the 25°C value. If not, raise
or lower the tempco and measure again until the 25°C value is
read.
C. Finding the TDS Ratio of an Unknown Solution
Once the effect of temperature is removed, the compensated
conductivity is a function of the concentration (TDS). There is a ratio
of TDS to compensated conductivity for any solution, which varies with
concentration. The ratio is set during calibration in USER mode as in
section User Programmable Conductivity to TDS Ratio, pg. 13. A truly
unknown solution has to have its TDS determined by evaporation and
weighing. Then the solution whose TDS is now known can be measured
for conductivity and the ratio calculated. Next time the same solution is
to be measured, the ratio is known.