Myron L 6P and 4P User Manual
Page 43
39
D. A Chart of Comparative Error
In the range of 1000 µS, the error using KCl on a solution that should be
compensated as NaCl or as 442, is illustrated in the graph below.
Users wanting to measure natural water based solutions to 1% would
have to alter the internal compensation to the more suitable preloaded
“442” values or stay close to 25°C. Users who have standardized to KCl-
based compensation may want to stick with it, regardless of increasing
error as you get further from 25°C. The Ultrameter II will provide the
repeatability and convertibility of data necessary for relative values for
process control.
E. Other Solutions
A salt solution like sea water or liquid fertilizer acts like NaCl. An internal
correction for NaCl can be selected for greatest accuracy with such
solutions. Many solutions are not at all similar to KCl, NaCl or 442. A
sugar solution, or a silicate, or a calcium salt at a high or low temperature
may require a “User” value peculiar to the application to provide readings
close to the true compensated conductivity.
Clearly, the solution characteristics should be chosen to truly represent
the actual water under test for rated accuracy of ±1%. Many industrial
applications have historically used relative measurements seeking a
7%
Chart 2
55
(1)%
(2)%
0%
1%
2%
3%
4%
5%
6%
0 5 10 15 20 25 30 35 40 45 50
Temperature
NaCl error with KCl tempco
442 error with KCl tempco