Myron L 6P and 4P User Manual
Page 17
13
2. Disabling Temperature Compensation
a.
Select USER mode, see Procedure to Select a Solution, pg. 11.
b.
With “USER” selected, press
. If the display does not
show .00%/°C, hold
long enough to bring the tempco to
.00%/°C (see Figure 3).
c.
Press
twice
(3 times if in TDS mode).
Temperature compensation
is now disabled (=0) for
measurements in USER mode.
3. User Programmable Conductivity to TDS Ratio
This feature allows you to select a custom conductivity to TDS conversion
ratio within the range of 0.20-7.99 for USER mode measurements.
To determine the conversion ratio for a custom solution of known TDS ppm
value, measure the solution conductivity at 25ºC with the Ultrameter II and
divide the ppm value by the µS value. For example, a solution of known 75ppm
TDS and measured 100µS conductivity at 25ºC would have a conversion
ratio of 75/100 or 0.75. Enter the new conversion ratio as follows:
a.
While in “USER” mode, press
.
b.
Press
twice (to skip over tempco adjustment) and
“RATIO” will appear (see
Figure 4).
c.
Adjust with
or
key until new conversion
ratio is displayed.
d.
Press
twice (to skip over calibration adjustment) to
accept new conversion ratio. You are now ready to measure
samples with the new conductivity/TDS ratio.
Figure 3
°C
% /
User
COND
Figure 4
RATIO
User
TDS
In these first six sections, you have learned
all you need to make accurate measurements.
The following sections contain calibration,
advanced operations and technical information.