Myron L 6P and 4P User Manual
Page 31
27
2.
Press
to display the stored memory records.
3.
Press
repeatedly until you pass the CELL ch location.
The display will show “Auto oFF” (see Figure 24).
4.
Press
to initiate. CAL
will be displayed along with
the “15 SEC” (see Figure
25).
5.
Press
or
to
change the time (see
Figure 26). Maximum
time is shown.
6.
Press
to accept the
change (new time).
XV.
USER MODE CALIBRATION LINC™ FUNCTION
Linc™ function allows easy calibration when in User mode and the user
does not have a user standard solution to calibrate the instrument. This
function will ensure more repeatable and accurate measurements than
many other calibration methods. It is recommended that this function be
used to provide the highest degree of confidence when the Ultrameter II
is used in “User” mode. When Linc is used, the User mode is linked to
another standard, i.e. if User and KCl are linked, a KCI standard solution
is used to calibrate the instrument. It is that simple.
A. Calibration of Ultrameter II for use in User Mode
1.
Press
or
key.
2.
Calibrate the unit using a Standard Solution, (ref.
CALIBRATION, pg. 14).
3.
Place the Ultrameter II in User mode, (ref. SOLUTION
SELECTION, pg. 11).
Figure 25
CAL
Figure 26
CAL