Myron L 6P and 4P User Manual
Page 16
12
you wish to change the
solution type.
2.
Press and hold
key
for 3 seconds to make
“SEL” appear (see Figure
1). For demonstration
purposes, all 4 solution
types are shown simultaneously.
3.
Use the
or
key to select type of solution desired
(ref. Solution Characteristics, pg. 40). The selected solution
type will be displayed: KCl, NaCl, 442 or User.
4.
Press
to accept new solution type.
E. Application of USER Solution Type
1. User Programmable Temperature Compensation
(Tempco)
This feature allows you to change your Ultrameter II’s temperature
compensating factor to another factor between 0-9.99%/°C (ref.
Temperature Compensation, pg. 37). This feature does not apply to pH
or ORP.
a.
As in Procedure to Select a Solution, pg. 11, select “USER”
mode.
b.
With “USER” mode now selected, press
. You may now
adjust a temperature compensation from .00%/°C to 9.99%/°C,
by pressing
or
.
See example in Figure 2.
c.
Press
twice to skip
calibration adjustment and
accept the new tempco (3
times if in TDS mode). You
are now ready to measure
samples with your new temperature compensation factor.
Figure 1
KCl
442
NaCl
User
Figure 2
°C
% /
User
COND