No response from datalogger through sc32b or modem, Peripheral – Campbell Scientific UT10 10 foot Tower w/ Grounding Kit and Hinged Base User Manual
Page 62

UT10 Weather Station
D. Disconnect any communications or storage peripherals from the
datalogger.
E. Reset the datalogger by turning the power switch to “OFF”, then to “ON”
or disconnecting and reconnecting the battery.
F. If still no response, call Campbell Scientific.
9.2.2 No Response from Datalogger through SC32B or Modem Peripheral
At the datalogger:
A. Make sure the battery has been installed, and the power switch, if any, is
“ON”.
B. Use a voltmeter to measure the voltage on the 12 V and G terminals; the
voltage must be between 9.6 and 16 Vdc.
C. Make sure the datalogger is connected to the modem, and the modem is
properly configured and cabled (Section 8.3, Communication and Data
Storage Peripherals).
At the computer:
D. Make sure the Station File is configured correctly (LoggerNet or PC400
Manual).
E. Check the cable(s) between the serial port and the modem. If cables have
not been purchased through Campbell Scientific, check for the following
configuration using an ohm meter:
25-pin serial port:
computer end modem end
2
2
3
3
7
7
20
20
9-pin serial port:
computer end modem end
2
3
3
2
4
20
5
7
F. Make sure the modem is properly configured and cabled (Section 8.3,
Communication and Data Storage Peripherals).
G. If still no response, call Campbell Scientific.
52