Section 7. Operation

configuration options. These are discussed below within the context of setting
up the sensor using a terminal emulator.

7.3 Terminal Mode

In normal operating mode the sensor will respond to a number of “root”
commands, shown in Table 7-1. Those commands allow the configuration of
the sensor in a non-interactive fashion, the polling of data and also switching
the sensor into an interactive user menu mode.

They are accessed via the command TERM PWS_1d Password

Commands are entered in the format ‘Command Pws_1d Password

TABLE 7-1. Command Set

Command Description

PASSWORD Sets

user

password

HELP*

Gives a list of available commands

OPEN* Activate

user

menu

TERM

Enter command mode

CLOSE Exit

command

mode

MSEND* Poll

messages

MSET

Set message parameters

HDATA

Retrieve historical particle data from sensor

SETPARAM Set

weather

parameters

FUZZY DIAG* Extra particular information (see Section 7.5.2)

CONFIG View

system configuration

SETCONFIG

Set system configuration

LOADOS

Load a new operating system

DIAGSET

Set diagnostic test parameters

DIAG

Run diagnostic test

TIME

Display or set time and date

RSENSOR

Receive remote sensor values (T °C and H %)

RESET Reset

hardware

XHMDATA

Historic `MSET’ data download using Xmodem

HMDATA

Output historic data to the command line

RUNDISC

Rotate the calibration disc for test / demonstration
purposes


*These commands are available outside terminal mode.

The most commonly used commands are MSEND, HELP and OPEN.

7-2