beautypg.com

Maintenance, Troubleshooting – Campbell Scientific LP02 Hukseflux Pyranometer User Manual

Page 22

background image

LP02 Pyranometer

8. Maintenance

Inspect and clean the outer dome at regular intervals, for example, every week

or so. Clean any accumulated dust, etc. off the dome and pyranometer body

using a soft cloth dampened with water or alcohol. Check that there is no

condensation within the dome.

It is also important to check the data returned from the sensor as it will show

the first indication of a fault. Be aware of several expected phenomena that can

cause strange measurements. In particular, on clear, windless nights the outer

dome temperature of horizontally placed pyranometers can fall as low as the

dewpoint temperature of the air, due to infrared radiation exchange with the

cold sky. (The effective sky temperature can be 30°C lower than the ground

temperature, which results in an infrared emission of –150 W m

–2

). If this

happens, dew, glazed frost or hoar frost can be precipitated on the top of the

outer dome and can stay there for several hours in the morning. An ice cap on

the dome is a strong diffuser and can decrease the pyranometer signal by up to

50% in the first hours after sunrise.

The calibration of the LP02 may drift with time and exposure to radiation.

Recalibration every two years is recommended. The sensor should be returned

to Campbell Scientific, the manufacturer, or a calibration lab with facilities to

calibrate radiation sensors.

9. Troubleshooting

Symptom: NAN, –9999, or radiation values around 0

1. Check that the sensor is wired to the differential channel specified by the

measurement instruction.

2. Verify that the range code is correct for the datalogger type.

3. Measure the impedance across the sensor wires. This should be around

100 Ω plus the cable resistance (typically 0.1 Ω m

–1

). If the resistance is

very low, there may be a short circuit (check the wiring). Resistances

somewhat lower than expected could be due to water ingress into the

sensor or enclosure connectors. If the resistance is infinite, there is a

broken connection (check the wiring).

4. Disconnect the sensor cable and check the voltage between pins 1 and 3

on the sensor. With the sensor located 8 inches below a 60 W

incandescent light bulb the voltage should be approximately 2.5 mV. No

voltage indicates a problem with the sensor.

Symptom: sensor signal is unrealistically high or low

1. Check that the right calibration factor has been properly entered into the

datalogger program. Please note that each sensor has its own individual

calibration factor.

2. Check the condition of the sensor cable.

16