HS2 and HS2P

Sensor rod length

20 cm and 12 cm
(NOT interchangeable)

20 cm and 12 cm
(interchangeable)

Accuracy

20 cm rods: ±3%

(EC ≤ 4 dS/m)

12 cm rods: ±3% (EC≤

6.5 dS/m)

±3% VWC

(EC < 2 dS/m)

GPS and geotagging

Yes

No

GPS zones

Yes

No

Data storage

Yes (1000+ readings)

No

Bluetooth

Yes

No

Date and time

Yes (with GPS accuracy

when available)

No

Display

128 x 64 pixel graphic

LCD

2 line x 16 char

alphanumeric LCD

LCD backlight

Yes

No

Firmware updates

User updateable

Factory only

Field carry case

Yes

No

Portable

Yes

Yes

Battery life

1000+ readings

2000+ readings

Unlike the HydroSense (I), the 12 cm and 20 cm rods of the

HydroSense II are not interchangeable.

The HydroSense II can display relative water content (RWC) based on wet and

dry references set by the user. The unit also displays water deficit, which

indicates how much water is required (in mm of applied water) to return the

soil to the previously saved wet reference.

Measurements are made by fully inserting the sensor rods into the soil and
pressing

. The process takes 2 to 3 seconds. The measurement can be

stored and later downloaded to a computer for display and analysis.

Communications between the computer and HydroSense II are via

Bluetooth

(p. 18). The HydroSense II includes a

GPS

(Global Positioning System)

(p. 19) receiver that enables data to be stored with position information

(

geotagged

(p. 19)).

The HydroSense II has these default power saving features:

turns off when idle for a configurable period

adjustable backlight brightness and on-time

backlight automatically disabled in bright conditions

GPS and Bluetooth can each be forced off.

NOTE