B.1.2. edlog single-ended measurement program, B.1.2 – Campbell Scientific HMP155A Temperature and Relative Humidity Probe User Manual
Page 30

Appendix B. Example Programs
B.1.2. Edlog Single-Ended Measurement Program
Below is an example CR10X program. Information about using it with other
Edlog dataloggers are provided in the program’s comments.
;Turn the HMP155A on.
;
01: Do (P86)
1: 41
Set Port 1 High
;Jumper wire from SW12V control to C1
;Green wire (C1) if using SW12V Power Switch
;For CR23X or CR5000 use 49 for SW12V internal
;control port
;Pause 2 seconds before making measurements so the
;probe can stabilize on true readings.
;
02: Excitation with Delay (P22)
1: 1
Ex Channel
2: 0
Delay W/Ex (units = 0.01 sec)
3: 500
Delay After Ex (units = 0.01 sec)
4: 0
mV Excitation
;Measure the HMP155A temperature.
;
03: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
;CR510, CR500 (2500mv); CR23X (1000 mV);
21X, CR7 (5000 mV)
3: 3
SE Channel
;Yellow wire (SE 3), white or purple wire (AG)
4: 1
Loc [ T_C ]
5: .14
Mult
;See TABLE 7-4 for alternative multipliers
6: -80
Offset
;See TABLE 7-4 for alternative offsets
;Measure the HMP155A relative humidity.
;
04: Volt (SE) (P1)
1: 1
Reps
2: 5
2500 mV Slow Range
;CR510, CR500 (2500 mV); CR23X (1000 mV);
21X, CR7 (5000 mV)
3: 4
SE Channel
;Blue wire (SE 4), white or purple wire (AG)
4: 2
Loc [ RH_pct ]
5: .1
Mult
;See TABLE 7-5 for alternative multipliers
6: 0
Offset
;Turn the HMP155A off.
;
05: Do (P86)
1: 51
Set Port 1 Low
;Jumper wire from SW12V control to C1
;Orange wire (C1) if older wiring
;Green wire (C1) if using SW12V device
;For CR23X or CR5000 use 59 for SW12V internal
;control port
B-2