beautypg.com

4 long lead lengths, Long lead lengths, 4. parameters for temperature – Campbell Scientific HMP155A Temperature and Relative Humidity Probe User Manual

Page 22: 5. parameters for relative humidity

background image

HMP155ATemperature and Relative Humidity Probe

datalogger control between measurements. For lead lengths greater than 6.1 m

(20 ft.) or when the probe will be continuously powered, use a differential

analog measurement. For a discussion on errors caused by long lead lengths

see Section 7.4.

HMP155A sensors purchased directly from Vaisala with serial

numbers < E4430001 require approximately 5 seconds warm up

time.

The HMP155A output scale is 0 to 1000 millivolts for the temperature range of

–80° to +60°C –112 to +140°F) and for the relative humidity range of 0 to

100%. Multipliers and offsets for converting voltage to temperature and

relative humidity are listed in TABLE 7-4 and TABLE 7-5 respectively.

TABLE 7-4. Parameters for Temperature

Units

Multiplier

(degrees mV

–1

)

Offset

(degrees)

Celsius

0.14

–80

Fahrenheit

0.252

–112

TABLE 7-5. Parameters for Relative Humidity

Units

Multiplier

(% mV

–1

)

Offset

(%)

Percent

0.1

0

Fraction

0.001

0

7.4 Long Lead Lengths

This section describes the error associated with measuring the HMP155A with

a single-ended measurement if the probe has a long cable. To avoid these

problems, Campbell Scientific recommends measuring the HMP155A using a

differential analog measurement (Instruction 2) when long lead lengths are

required. Generic datalogger connections for measuring the HMP155A using a

differential measurement are given in TABLE 7-2.

Understanding the details in this section is not required for the general

operation of the HMP155A with Campbell Scientific’s dataloggers.

The signal reference (white) and the power ground (black) are in common

inside the HMP155A. When the HMP155A temperature and relative humidity

are measured using a single-ended analog measurement, both the signal

reference and the power ground are connected to ground at the datalogger. The

signal reference and the power ground both serve as the return path for 12 V.

There will be a voltage drop along those leads because the wire itself has

resistance. The HMP155A draws approximately 4 mA when it is powered.

The wire used in the HMP155A (pn 9721) has resistance of 27.7 Ω/1000 feet.

Since the signal reference and the power ground are both connected to ground

at the datalogger, the effective resistance of those wires together is half of 27.7

NOTE

14