Campbell Scientific CSIM11 pH and ORP Sensors User Manual
Page 23

CSIM11 pH and ORP Probes
sensor in 3.5 molar KCl, or water if KCl is not available, and heat to
approximately 60 °C for 15 minutes; retest.
SYMPTOM: Slow response and/or noisy, erratic readings.
POSSIBLE REASON: Slow response is caused by a very high impedance in
either glass or reference electrode.
CHECK:
1. Visually inspect pH bulb and reference junction for coating or clogging. If
coated, clean as described in Section 9.2, Electrode Cleaning.
2. Allowing a pH electrode to dry out raises the impedance dramatically.
Soaking the electrode in 0.1 normal HCl for 1/2 hour and rinsing with
distilled water should speed response (always keep electrodes wet).
3. Chemical degradation of pH glass can occur rapidly in a high temperature
or high pH environment. This also increases impedance yielding sluggish
response.
4. Low temperature environments also increase impedance, with impedance
doubling every 8 °C temperature drop from 25 °C.
5. Test wiring for intermittent connections by moving wires at the connector
on the datalogger wiring panel. Tighten as necessary. This is a common
source of erratic signals.
SYMPTOM: No response to pH change.
POSSIBLE REASON: A cracked glass bulb.
CHECK:
If electrode reads between 5.8 and 6.2 pH in all solutions, visually inspect glass
bulb for cracks. Discard probe if it is cracked.
POSSIBLE REASON: Short circuit.
CHECK:
If electrode reads 7.0 pH or 0.0 mV when connected to the datalogger, visually
inspect cable for damage. If there is no visible damage, test electrode in 4.0
buffer.
15