beautypg.com

Troubleshooting – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 22

background image

CSIM11 pH and ORP Probes

10 minutes each. After treating the electrode with these strong acids or bases,

rinse the electrode with water and soak it in pH 4 buffer for at least 1/2 hour.

Greasy and oily coatings are best removed with a detergent solution or a

solvent that will not attack the electrode’s body. Methanol and isopropyl

alcohol are good choices for solvents. A soft toothbrush can be used with the

detergent to remove stubborn coatings.

Acetone, MEK, THF, or trichloroethane will irreparably

harm the electrode.

The pores of the reference junction may become clogged and surface cleaning

may not restore proper function. The electrode should then be heated to 60 °C

in 3 molar KCl and allowed to cool in the same solution. Rinse it with distilled

water and soak in pH 4 buffer for 1/2 hour before testing. The electrolyte

should be removed and replaced with fresh electrolyte before treating as above.

Electrodes age with time and eventually become desensitized. Extended

periods of service at temperatures greater than 80 °C or exposure to deionized

water accelerate this phenomenon. As a last resort, dip the electrode in a 10

percent ammonium bifluoride solution for 10 to 20 seconds, then rinse it with

tap water and soak it in 5-6 molar HCl for 30 seconds. Rinse it with tap water

and soak it in pH 4 buffer for 1/2 hour before testing.

The platinum sensing tip of an ORP electrode should be cleaned just like a pH

electrode. The surface can also be cleaned with an abrasive as a last resort.

Gently scour the platinum with a 600 grit wet emery cloth or preferably 1-3

micron alumina polishing powder.

10. Troubleshooting

SYMPTOM: Probe pegs at 14 pH or drifts off scale high.

POSSIBLE REASON: Open circuit in either glass electrode or reference

electrode.

CHECK:

1. Visually inspect cable and connector looking for a crushed or broken cable

jacket or a brittle cable jacket due to exposure to solar radiation. Discard

electrode if damage is present.

2. Move wires at datalogger to test for intermittent connection. Tighten

connectors if necessary.

3. Visually inspect bulb for a coating. If coated, use an appropriate solvent

or a high-quality detergent with a cotton swab to wipe bulb clean. Rinse

well with distilled water; soak in 4.0 buffer for at least 10 minutes, retest.

If electrode now responds, but erratically, soak in 10 percent HCl solution

for five (5) minutes, rinse well with distilled water; soak in 4.0 buffer for

at least 10 minutes, retest.

4. Visually inspect reference junction (large white surface at front of sensor).

If coated, clean by rinsing well; retest. If electrode still reads high, place

WARNING

14