5. Under the
Available Sensors and Devices list, select the Sensors |
Temperature folder. Select 107 Temperature Probe. Click
to
move the selection to the
Selected device window. Use the data default of
degree Celsius.
6. Under the
Available Sensors and Devices list, select the Sensors | Water |
Quality folder. Select CSIM11 pH Probe. Click
to move the
selection to the
Selected device window. Click on the Solution
temperature (Deg C) reference box and select T107_C. The probe
usually requires an offset adjustment. The white panel at the bottom of the
Properties window provides a procedure for determining the value that
should be entered in the
Offset box.