beautypg.com

Calibration/orp check, 1 ph sensor calibration, 2 orp check – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 20: Maintenance, Ph sensor calibration, Orp check

background image

CSIM11 pH and ORP Probes

8. Calibration/ORP Check

8.1 pH Sensor Calibration

Calibration should be carried out according to the detailed procedure later in

this document (see Appendix C, Detailed Calibration Procedure and

Manufacturer Tips). The following paragraphs are for general information.

The calibration should use two or more pH standards. It is recommended that

pH 7 buffer be used to check the zero point and at least an acid or alkaline

buffer, that brackets the sample’s pH value, be used to set the slope.

An electrode measuring many samples a day should be calibrated at least once

a day. The frequency of calibration will depend on the level of accuracy

required and the coating/fouling nature of the samples being measured.

Electrodes that are continuously monitoring a sample should be checked at

least once a week or whatever period experience dictates.

Grab Sample Calibration is a technique where the process electrode has been

calibrated and placed on line for some period of time. Its output is then

verified by measuring the pH of a sample with another electrode. The grab

sample should be measured as soon as possible to avoid errors caused by

changes in the sample’s temperature or changes in the samples pH due to

exposure to the atmosphere.

8.2 ORP Check

Check the ORP sensor when it’s initially deployed and after three months of

field service. To check the sensor, place it in a known millivolt solution. The

sensor manufacturer offers +230 mV and +470 mV solutions. If the sensor

reading is within ±20 mV of the millivolt value of the solution, the sensor is

operating properly.

9. Maintenance

Developing an effective maintenance schedule is incumbent on understanding

the process effects that are specific to your application. A pH sensor develops

a millivolt potential directly proportional to the free hydrogen ion

concentration in an aqueous solution. The sensor is composed of a reference

electrode and its gelled reference electrolyte, a measurement electrode exposed

to the process solution, and a porous junction that maintains electrical contact

between the two. Porous PTFE is the newest

technology in reference

junctions. Wedgewood Analytical, Inc. offers a patented porous PTFE liquid

junction which is chemically inert; and is chemically compatible with virtually

all chemicals.

High Temperature which causes 1) Faster Response / Lower Impedance;

2) Aging acceleration, Lithium Ions Leached from Membrane; 3) Short

Span

Low Temperature which causes Slower Response / Higher Impedance

Measurement > 10.0 pH causes Alkaline / Sodium Ion Error

12