beautypg.com

1 replacing reference electrolyte, 2 electrode cleaning, Replacing reference electrolyte – Campbell Scientific CSIM11 pH and ORP Sensors User Manual

Page 21: Electrode cleaning

background image

CSIM11 pH and ORP Probes

Coatings can cause 1) Slower Response; 2) Zero Offset increase;

3) Dehydration

Steam Sterilization causes 1) Dehydration; 2) Ag/AgCl dissolving from

Silver Reference Element

9.1 Replacing Reference Electrolyte

Readings that drift indicate the electrolyte should be checked.

Refill Procedure:

1. Clean the probe tip as discussed in the maintenance procedures below.

2. Remove the reference reservoir plug. Place it where it will stay clean.

3. Rinse the reservoir with deionized water repeatedly to remove the old

solution. Drain out all remaining water.

4. Completely fill the reservoir with the new reference solution. It does not

take very much. Make sure you keep the bottle tip clean, and replace the

bottle cap immediately after using the bottle.

The bottle of refill solution contains undissolved salts; this is to

ensure the solution remains saturated.

5. Replace the red plug applying new pipe tape. Make sure the plug is

screwed back in as far as it was originally.

Because the new reference solution is viscous, you may need to wait a few

minutes while the reference solution settles. If necessary, take the probe in

hand and gently swing it in a downward arc to speed the flow of solution and

remove air pockets. Add more solution as needed to completely fill the

reservoir.

9.2 Electrode Cleaning

pH and ORP sensors require more maintenance than many other types of

sensors. The ORP platinum band can foul with algae and other biological

sources. Cleaning will be required approximately every 1 to 2 weeks. Fouling

can be minimized by locating the probe in a very dark place.

Slow response and large offsets may indicate the measurement electrode has

become coated. The nature of the coating will dictate the type of cleaning

technique that should be used.

Soft coatings, like foodstuffs or bacterial films are best removed using a squirt

bottle or the water jet from a faucet. If this is not successful, then wiping with

a soft wet cloth is the best choice.

Hard coatings, like calcium or lime scale are best removed with a solvent

appropriate for the particular coating. A 5 percent solution of HCl would be a

good choice for the calcium scale. If unsure of the proper solvent to remove a

hard mineral coating, alternate between 5 percent HCI and 4 percent NaOH for

NOTE

13