Appendix A. CS700H Operation

Factory settings have been set to adequately measure
precipitation during cold precipitation events. Changing
these settings is not recommended, and doing so may
change the data outcome or render the sensor inoperable.

CAUTION

Table A-1 shows the CS700H’s factory defaults:

TABLE A-1. CS700H Default Values

External Control Off (=> Auto)

(X20 Command) = 0

Snow Sensor Enabled

(X22 Command) = 1

Active On Temperature +4°C

(X23 Command)

Active Off Temperature+5°C

(X24 Command)

Low Off Temperature –20°C

(X25 Command)

Funnel Set Point Temp +10°C

(X26 Command)

Snow Run-On Time 18 mins

(X27 Command)

Units °C

(X28 Command) = 0

FIGURE A-1. Locations of the CS700H’s Heater Components

A-1