CS700 and CS700H Rain Gage

8. Calibration

The sensor is factory calibrated; recalibration is not required unless damage
has occurred or the adjustment screws have loosened.

Nevertheless, the following calibration check is recommended once every 12
months:

a. Remove the housing assembly from the base by removing the three

screws and lifting upward on the housing.

b. Check the bubble level to verify the rain gage is level.

c. Pour water through the inner funnel to wet the two bucket surfaces.

Using a graduated cylinder, slowly pour 314 cc (19.16 in

3

) of water, over

a 15 minute period, into the collection funnel. This volume of water is
equal to .39 in of rainfall (10 mm).

d. After the water has passed through the rain gage, the tipping bucket

should have tipped 39 times.

e. If the rain gage fails to record the correct number of tips, return the unit to

Campbell Scientific for recalibration.

Factory Calibration

If factory calibration is required, contact Campbell Scientific to obtain an
RMA (see Warranty and Assistance in the front of the manual).