V (eq. 4). the estimated measurement accuracy of m – Campbell Scientific CS110 Electric Field Meter User Manual

Page 21

background image

CS110 Electric Field Meter

of variations in surface cleanliness along with charging and discharging
processes. Polished 316-L stainless-steel is used for critical electrode surfaces
on the CS110 to minimize unwanted surface charges. CS110s with clean
electrodes have been found to display electric field offsets

<⏐30 V/m⏐, which

has negligible effect on the determination of M

parallel_plate

because of the

±15

kV/m range of electric fields used during factory calibration. Neglecting
O

parallel_plate

results in the simplified parallel-plate calibration equation

E

=

M

parallel_plate

⋅V (eq.

4).

The estimated measurement accuracy of M

parallel_plate

for the CS110 calibrated in

the parallel plate electric field calibrator illustrated in Figure 5 is

± 1 %. The

electric field offset of the CS110 can be measured by covering the stator with a
clean Zero Electric Field Cover (PN: 17642). If the resulting zero field reading
with the zero field cover exceeds an absolute value of 60 V/m then cleaning of
electrodes in the CS110 is suggested. The factory calibration data for a typical
CS110 factory calibration and resulting determination of M

parallel_plate

= 84.32

V/m

⋅mV (Volts/meter⋅millivolt) is illustrated in Figure 6.

y = 84.324x + 26.258

R

2

= 1

-15000

-10000

-5000

0

5000

10000

15000

20000

-200

-150

-100

-50

0

50

100

150

200

250

A

pplie

d Ele

c

tr

ic

Fie

ld (V/m

)

-20000

Charge Amplifier Output Voltage (mV)

FIGURE 6. Factory calibration data for CS110 SN: 1026

13