Appendix A. Theory of Operation

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

0

-20

-40

-60

-80

20

40

60

80

Samples

mV

1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

1250

1200

1150

1100

1300

1350

1400

Samples

mV

1250 mV Offset (needed for CR200X)

0 mV Offset for Dataloggers other than CR200X

FIGURE A-8. Adding 1250 mV creates positive output

1000

500

0

1500

2000

2500

mV

Adjusted Measurement Range

mV signal CR200X

FIGURE A-9. CS11-L measurement range

A.6 Multiple Passes Through the Sensor

Multiple passes can pass through the sensor to amplify the signal of the
amperage being measured (FIGURE A-10). However, the multiplier will need
to be changed, depending on how many passes through the sensor.

The range code needs to be changed to match the number of wire
passes through the sensor.

NOTE

Passes New

Multiplier Range

Code

2 0.1

x2

4 0.05

x4

5 0.04

x5

8 0.025

x8

10 0.02

x10

20 0.01

x20