CMP6-L, CMP11-L, and CMP21-L Pyranometers

TABLE 7-2. CMP6 and CMP11 Single-Ended Connections to Campbell Scientific Dataloggers



Color



Description

CR9000(X), CR5000,
CR3000, CR1000,
CR800


CR510, CR500,
CR10(X)



21X, CR7, CR23X

White

Signal (+)

SE Analog

SE Analog

SE Analog

Black Signal

(–)

AG

Clear Shield

G

7.2.3 CMP21 Wiring

A CMP21 purchased from Campbell Scientific has different
wiring than a CMP21 purchased directly from Kipp & Zonen.

NOTE

The CMP21’s pyranometer can be measured using either differential analog
channels or single-ended analog channels. A differential voltage measurement
is recommended because it has better noise rejection than a single-ended
measurement. If a differential channel is not available, a single-ended
measurement can be used.

A single-ended channel and a voltage excitation channel are used to measure
the CMP21’s internal thermistor.

Connections to Campbell Scientific dataloggers for a differential measurement
are given in TABLE 7-3. A user-supplied jumper wire should be connected
between the low side of the differential input and ground (AG or

) to keep

the signal in common mode range. Connections to Campbell Scientific
dataloggers for a single-ended measurement are given in TABLE 7-4.

TABLE 7-3. CMP21 Differential Connections to Campbell Scientific Dataloggers



Wire Color


Wire Label/
Description

CR9000(X),
CR5000, CR3000,
CR1000, CR800


CR510, CR500,
CR10(X)



21X, CR7, CR23X

White

Pyranometer Sig

DIFF Analog High

DIFF Analog High

DIFF Analog High

Blue

Pyranometer Ref

*DIFF Analog Low

*DIFF Analog Low

*DIFF Analog Low

Yellow

Thermistor Volt Excite

VX or EX

E

EX

Black

Thermistor Sig

Single-ended analog

Single-ended analog

Single-ended analog

Brown Thermistor

Ref

AG

Clear Shield

G


* Jumper to AG or

with user-supplied wire.

13