1 cmp6, cmp11, and cmp21 thermopile schematic, 2 cmp6 and cmp11 wiring, Cmp6, cmp11, and cmp21 thermopile schematic – Campbell Scientific CMP6-L, CMP11-L, and CMP21-L Pyranometers User Manual
Page 20: Cmp6 and cmp11 wiring, Scientific dataloggers

CMP6-L, CMP11-L, and CMP21-L Pyranometers
7.2.1 CMP6, CMP11, and CMP21 Thermopile Schematic
A schematic diagram of a CMP6, CMP11, or CMP21 thermopile is shown in
FIGURE 7-4.
Red
Blue
Black
FIGURE 7-4. CMP6, CMP11, and CMP21 thermopile detector
schematic
W
hite (+)
Black (-)
Shield
7.2.2 CMP6 and CMP11 Wiring
A CMP6 or CMP11 purchased from Campbell Scientific has
different wiring than a pyranometer purchased directly from
Kipp & Zonen.
NOTE
The pyranometer is measured using either differential analog channels or
single-ended analog channels.
A differential voltage measurement is recommended because it has better noise
rejection than a single-ended measurement.
Connections to Campbell Scientific dataloggers for a differential measurement
are given in TABLE 7-1. A user-supplied jumper wire should be connected
between the low side of the differential input and ground (AG or
) to keep
the signal in common mode range.
Connections to Campbell Scientific dataloggers for a single-ended
measurement are given in TABLE 7-2.
TABLE 7-1. CMP6 and CMP11 Differential Connections to Campbell Scientific Dataloggers
Color
Description
CR9000(X), CR5000,
CR3000, CR1000,
CR800
CR510, CR500,
CR10(X)
21X, CR7, CR23X
White
Signal (+)
DIFF Analog High
DIFF Analog High
DIFF Analog High
Black
Signal (–)
*DIFF Analog Low
*DIFF Analog Low
*DIFF Analog Low
Shield Shield
G
* Jumper to AG or
with user supplied 26 AWG or larger wire.
12